EYEMATE-SC Trial: Twelve-Month Safety, Performance, and Accuracy of a Suprachoroidal Sensor for Telemetric Measurement of Intraocular Pressure

Ophthalmology. 2023 Mar;130(3):304-312. doi: 10.1016/j.ophtha.2022.09.021. Epub 2022 Oct 3.

Abstract

Purpose: Measuring and controlling intraocular pressure (IOP) provide the foundation for glaucoma treatment. Self-tonometry has been proposed as an alternative to measure IOP throughout the entire day better. The novel EYEMATE-SC sensor (Implandata) is implanted in the suprachoroidal space to enable contactless continual IOP monitoring. The aim of the present study was to investigate the 1-year safety, performance, and accuracy of the EYEMATE-SC in patients with primary open-angle glaucoma undergoing simultaneous nonpenetrating glaucoma surgery (NPGS).

Design: Prospective, multicenter, open-label, single-arm, interventional clinical trial.

Participants: Twenty-four eyes of 24 patients with primary open-angle glaucoma who were due to undergo NPGS (canaloplasty or deep sclerectomy).

Methods: An EYEMATE-SC sensor was implanted during NPGS. Goldmann applanation tonometry (GAT) measurements were compared with the sensors' IOP measurements at all postoperative visits through 12 months.

Main outcome measures: Device position and adverse events.

Results: Fifteen eyes underwent canaloplasty, and 9 underwent deep sclerectomy. Successful implantation of the sensor was achieved in all eyes with no reported intraoperative difficulties. Through the 12-month follow-up, no device migration, dislocation, or serious device-related complications were recorded. A total of 536 EYEMATE-SC measurements were pairwise included in the IOP agreement analysis. The overall mean difference between GAT and EYEMATE-SC measurements was 0.8 mmHg (95% confidence interval [CI] of the limits of agreement [LoA], -5.1 to 6.7 mmHg). The agreement gradually improved, and from 3 months after surgery until the end of the follow-up, the mean difference was -0.2 mmHg (95% CI of LoA, -4.6 to 4.2 mmHg) over a total of 264 EYEMATE-SC measurements, and 100% of measurements were within ±5 mmHg of GAT.

Conclusions: The EYEMATE-SC sensor was safe and well tolerated through 12 months. Moreover, it allowed accurate, continuous IOP monitoring.

Financial disclosure(s): Proprietary or commercial disclosure may be found after the references.

Keywords: Glaucoma; IOP; Implantable; Self-tonometry; Telemetry.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Glaucoma, Open-Angle*
  • Humans
  • Intraocular Pressure*
  • Prospective Studies
  • Reproducibility of Results
  • Tonometry, Ocular