Radiation dose and image quality of CT coronary angiography in patients with high heart rate or irregular heart rhythm using a 16-cm wide detector CT scanner

Medicine (Baltimore). 2022 Sep 16;101(37):e30583. doi: 10.1097/MD.0000000000030583.

Abstract

Aim of the study was to evaluate the effect of high and irregular heart rate on the image quality and on the radiation exposure using a 256-row, 16-cm wide detector computed tomography (CT) system. Between March and December 2019, 349 patients undergoing CT coronary angiography (CTCA) were prospectively enrolled. Patients were divided into 2 study groups; Group 1 included patients with a regular heart rate of ≤70 bpm, while Group 2 included patients with an irregular heart rhythm or heart rate of >70 bpm. In all patients, image quality score and radiation dose were analyzed and recorded. In Group 1, there were a total of 195 patients, while in Group 2, there were 154 patients. Of the 349 patients, 299 of them had a regular heart rhythm (85.7%) and 50 (14.3%) had an irregular heart rhythm. Mean heart rate during scanning was 59 ± 7 bpm in Group 1 and 80 ± 12 bpm in Group 2. Mean effective dose of CTCA in Group 1 (1.2 ± 0.8 mSv) was lower than in Group 2 (1.9 ± 1.2 mSv, P < .001). Mean image quality (Likert score) of Group 1 was significantly higher than in Group 2 (4.1 vs 3.4, P < .001). CT scanner with 16-cm wide detector enables low-radiation exposure during CTCA even at high heart rate or irregular heart rhythm. Good CTCA image quality and low dose are related to low heart rate.

MeSH terms

  • Coronary Angiography / methods
  • Heart Rate
  • Humans
  • Radiation Dosage
  • Tomography Scanners, X-Ray Computed
  • Tomography, X-Ray Computed*