The SIDECAR project: S-IcD registry in European paediatriC and young Adult patients with congenital heaRt defects

Europace. 2023 Feb 16;25(2):460-468. doi: 10.1093/europace/euac162.

Abstract

Aims: Subcutaneous-implantable cardiac defibrillators (S-ICDs) are used increasingly to prevent sudden cardiac death in young patients. This study was set up to gain insight in the indications for S-ICD, possible complications, and their predictors and follow-up results.

Methods and results: A multicentre, observational, retrospective, non-randomized, standard-of-care registry on S-ICD outcome in young patients with congenital heart diseases (CHDs), inherited arrhythmias (IAs), idiopathic ventricular fibrillation (IVF), and cardiomyopathies (CMPs). Anthropometry was registered as well as implantation technique, mid-term device-related complications, and incidence of appropriate/inappropriate shocks (IASs). Data are reported as median (interquartile range) or mean ± standard deviation. Eighty-one patients (47% CMPs, 20% CHD, 21% IVF, and 12% IA), aged 15 (14-17) years, with body mass index (BMI) 21.8 ± 3.8 kg/m2, underwent S-ICD implantation (primary prevention in 59%). This was performed with two-incision technique in 81% and with a subcutaneous pocket in 59%. Shock and conditional zones were programmed at 250 (200-250) and 210 (180-240) b.p.m., respectively. No intraoperative complications occurred. Follow up was 19 (6-35) months: no defibrillation failure occurred, 17% of patients received appropriate shocks, 13% of patients received IAS (supraventricular tachycardias 40%, T-wave oversensing 40%, and non-cardiac oversensing 20%). Reprogramming, proper drug therapy, and surgical revision avoided further IAS. Complications requiring surgical revision occurred in 9% of patients, with higher risks in patients with three-incision procedures [hazard ratio (HR) 4.3, 95% confidence interval (95% CI) 0.5-34, P = 0.038] and BMI < 20 (HR 5.1, 95% CI 1-24, P = 0.031).

Conclusion: This multicentre European paediatric registry showed good S-ICD efficacy and safety in young patients. Newer implantation techniques and BMI > 20 showed better outcome.

Keywords: Congenital heart disease; ICD complications; Implantable cardioverter defibrillator; Paediatric age; Subcutaneous-ICD.

MeSH terms

  • Arrhythmias, Cardiac / diagnosis
  • Arrhythmias, Cardiac / epidemiology
  • Arrhythmias, Cardiac / therapy
  • Child
  • Death, Sudden, Cardiac / epidemiology
  • Death, Sudden, Cardiac / etiology
  • Death, Sudden, Cardiac / prevention & control
  • Defibrillators, Implantable* / adverse effects
  • Heart Defects, Congenital* / complications
  • Heart Defects, Congenital* / diagnosis
  • Heart Defects, Congenital* / epidemiology
  • Humans
  • Registries
  • Retrospective Studies
  • Treatment Outcome
  • Young Adult

Supplementary concepts

  • Paroxysmal ventricular fibrillation