Design and calibration of a Tonpilz transducer for low frequency medical ultrasound tomography

Annu Int Conf IEEE Eng Med Biol Soc. 2022 Jul:2022:4611-4617. doi: 10.1109/EMBC48229.2022.9872007.

Abstract

The design and performance of a transducer for low frequency ultrasound tomography is presented, motivated by recent research demonstrating that acoustic waves transmitting at frequencies between 10 kHz and 750 kHz penetrate the lungs and may be useful for thoracic imaging. An adaptation of the traditional Tonpilz design was developed, vibrational amplitude and electrical impedance were measured, and an optimal frequency was determined. The design is found to meet the desired mechanical, electrical, and safety specifications. Thus, it was considered a promising option for the target application of pulmonary imaging with ultrasound computed tomography between 50 and 200 kHz; highest efficiency achieved around 125 kHz and 156 kHz, and beam divergence of 40°.