Inner branched complex aortic repair outcomes from a national multicenter registry using the E-xtra design platform

J Vasc Surg. 2023 Feb;77(2):338-346. doi: 10.1016/j.jvs.2022.08.034. Epub 2022 Sep 5.

Abstract

Background: Complex aortic pathology still represents an open issue in contemporary endovascular management, with continuous technological advancement being introduced in practice over time aiming to improve outcomes. Thus far, the dualism between the fenestrated and branched configuration for visceral artery revascularization is yet unsolved, with each approach having its own pros and cons. The inner branched technology for endovascular aneurysm repair (iBEVAR) aims to take the best out of both strategies, offering wide applicability and stable bridging stent sealing. The objective of this study was to evaluate the early outcomes obtained with a single manufacturer custom-made inner-branched endograft in a multicenter Italian experience.

Methods: All patients consecutively treated with E-xtra design devices in three Italian facilities were enrolled. Anatomic characteristics and perioperative data were analyzed. The main objective was to asses technical and clinical success after iBEVAR. Secondary end points were overall survival, aortic-related mortality, target visceral vessel (TVV) patency, and freedom from target vessel instability during follow-up.

Results: From 2016 to 2021, 45 patients were treated with an E-xtra design device revascularizing at least one visceral vessel through an inner branch. The mean age at the time of the procedure was 71.1 ± 9.3 years and 77.8% were males. The total number of target visceral arteries to be bridged with an inner branch was 159. The extent of aortic repair was thoracoabdominal in 91.1% of the cases. Technical success was achieved in 93.3% of the procedures (42/45) with all failures owing to a type I endoleak at final angiography. Each TVV was successfully connected to the graft's main body as planned without complications. Following their intervention, five patients developed spinal cord ischemia and in three of these cases symptoms persisted after discharge (6.7%). At 30 days clinical success was 93.3% (42/45). No death as well as no TVV thrombosis occurred within 30 days from the primary procedures. The mean follow-up was 22.8 ± 14.2 months. The Kaplan-Meier estimate of overall survival and TVV patency at 36 months were 83.9% and 95.9%, respectively.

Conclusions: Inner branches seem to be a promising technology in the complex aortic repair landscape, with an applicability ranging from type II thoracoabdominal aneurysm to type I endoleak repair after infrarenal endografting. Whether iBEVAR could offer results comparable with those provided by fenestrated/branched endovascular aneurysm repair in terms of target vessel patency and stent stability is yet to be established and further studies are, therefore, needed.

Keywords: Branched endograft; Complex aortic repair; Endovascular repair; Inner branches; Thoracoabdominal aneurysm.

Publication types

  • Multicenter Study

MeSH terms

  • Aortic Aneurysm, Abdominal* / complications
  • Aortic Aneurysm, Abdominal* / diagnostic imaging
  • Aortic Aneurysm, Abdominal* / surgery
  • Aortic Aneurysm, Thoracic* / complications
  • Aortic Aneurysm, Thoracic* / diagnostic imaging
  • Aortic Aneurysm, Thoracic* / surgery
  • Aortography / methods
  • Blood Vessel Prosthesis / adverse effects
  • Blood Vessel Prosthesis Implantation*
  • Endoleak / surgery
  • Endovascular Procedures*
  • Female
  • Humans
  • Male
  • Postoperative Complications
  • Prosthesis Design
  • Registries
  • Risk Factors
  • Time Factors
  • Treatment Outcome