Artificial intelligence based image quality enhancement in liver MRI: a quantitative and qualitative evaluation

Radiol Med. 2022 Oct;127(10):1098-1105. doi: 10.1007/s11547-022-01539-9. Epub 2022 Sep 7.

Abstract

Purpose: To compare liver MRI with AIR Recon Deep Learning™(ARDL) algorithm applied and turned-off (NON-DL) with conventional high-resolution acquisition (NAÏVE) sequences, in terms of quantitative and qualitative image analysis and scanning time.

Material and methods: This prospective study included fifty consecutive volunteers (31 female, mean age 55.5 ± 20 years) from September to November 2021. 1.5 T MRI was performed and included three sets of images: axial single-shot fast spin-echo (SSFSE) T2 images, diffusion-weighted images(DWI) and apparent diffusion coefficient(ADC) maps acquired with both ARDL and NAÏVE protocol; the NON-DL images, were also assessed. Two radiologists in consensus drew fixed regions of interest in liver parenchyma to calculate signal-to-noise-ratio (SNR) and contrast to-noise-ratio (CNR). Subjective image quality was assessed by two other radiologists independently with a five-point Likert scale. Acquisition time was recorded.

Results: SSFSE T2 objective analysis showed higher SNR and CNR for ARDL vs NAÏVE, ARDL vs NON-DL(all P < 0.013). Regarding DWI, no differences were found for SNR with ARDL vs NAÏVE and, ARDL vs NON-DL (all P > 0.2517).CNR was higher for ARDL vs NON-DL(P = 0.0170), whereas no differences were found between ARDL and NAÏVE(P = 1). No differences were observed for all three comparisons, in terms of SNR and CNR, for ADC maps (all P > 0.32). Qualitative analysis for all sequences showed better overall image quality for ARDL with lower truncation artifacts, higher sharpness and contrast (all P < 0.0070) with excellent inter-rater agreement (k ≥ 0.8143). Acquisition time was lower in ARDL sequences compared to NAÏVE (SSFSE T2 = 19.08 ± 2.5 s vs. 24.1 ± 2 s and DWI = 207.3 ± 54 s vs. 513.6 ± 98.6 s, all P < 0.0001).

Conclusion: ARDL applied on upper abdomen showed overall better image quality and reduced scanning time compared with NAÏVE protocol.

Keywords: Artificial intelligence; Image quality; Scanning time; Sequences optimization.

MeSH terms

  • Adult
  • Aged
  • Artificial Intelligence*
  • Diffusion Magnetic Resonance Imaging / methods
  • Echo-Planar Imaging* / methods
  • Female
  • Humans
  • Image Enhancement / methods
  • Liver / diagnostic imaging
  • Magnetic Resonance Imaging
  • Middle Aged
  • Prospective Studies
  • Reproducibility of Results