Wireless Measurements Using Electrical Impedance Spectroscopy to Monitor Fracture Healing

Sensors (Basel). 2022 Aug 19;22(16):6233. doi: 10.3390/s22166233.

Abstract

There is an unmet need for improved, clinically relevant methods to longitudinally quantify bone healing during fracture care. Here we develop a smart bone plate to wirelessly monitor healing utilizing electrical impedance spectroscopy (EIS) to provide real-time data on tissue composition within the fracture callus. To validate our technology, we created a 1-mm rabbit tibial defect and fixed the bone with a standard veterinary plate modified with a custom-designed housing that included two impedance sensors capable of wireless transmission. Impedance magnitude and phase measurements were transmitted every 48 h for up to 10 weeks. Bone healing was assessed by X-ray, µCT, and histology. Our results indicated the sensors successfully incorporated into the fracture callus and did not impede repair. Electrical impedance, resistance, and reactance increased steadily from weeks 3 to 7-corresponding to the transition from hematoma to cartilage to bone within the fracture gap-then plateaued as the bone began to consolidate. These three electrical readings significantly correlated with traditional measurements of bone healing and successfully distinguished between union and not-healed fractures, with the strongest relationship found with impedance magnitude. These results suggest that our EIS smart bone plate can provide continuous and highly sensitive quantitative tissue measurements throughout the course of fracture healing to better guide personalized clinical care.

Keywords: Bluetooth transmission; bone fracture repair; electrical impedance spectroscopy; longitudinal monitoring; smart bone plate; wireless measurements.

MeSH terms

  • Animals
  • Bone Plates
  • Bony Callus / diagnostic imaging
  • Bony Callus / pathology
  • Dielectric Spectroscopy / methods
  • Fracture Healing*
  • Fractures, Bone* / diagnostic imaging
  • Rabbits