Atlantoaxial Joint Distraction and Fusion with DTRAX Intra-Articular Cages: A Cadaveric Feasibility Study and Review of the Pertinent Literature

World Neurosurg. 2022 Oct:166:153-158. doi: 10.1016/j.wneu.2022.07.077. Epub 2022 Jul 30.

Abstract

Purpose: Atlantoaxial joint distraction is a key procedure for the treatment of selected patients affected by basilar invagination (BI). In recent years, several authors have reported various techniques of distraction and fixation of the C1-C2 joint using different types of intra-articular spacers, with or without posterior fixation. We review the pertinent literature and propose a feasibility study on the use of a new device for the distraction of the C1-C2 joint aimed to the descent of the dens out of the foramen magnum suggesting its application on selected cases of BI.

Methods: The GL-DTRAX Cervical Cage-SE is a cage approved by the Food and Drug Administration for distraction and fixation of subaxial cervical spine. Five adult cadaveric specimens were dissected surgically to evaluate the feasibility of DTRAX insertion inside the C1-C2 joint through a posterior approach.

Results: The cages were uneventfully set into the C1-C2 intra-articular space of all samples without the need to sacrifice C2 nerve roots and ganglia. Postoperative cervical computed tomography scanning confirmed the correct fitting of the devices in every sample.

Conclusions: This cadaveric study highlights the feasibility of the DTRAX cage as a C1-C2 intra-articular device producing a substantial distraction of atlantoaxial complex and suggesting a possible therapeutic role in selected cases of BI.

Keywords: Basilar invagination; DTRAX; Feasibility cadaveric study; Intra-articular C1–C2 spacers.

Publication types

  • Review

MeSH terms

  • Adult
  • Atlanto-Axial Joint* / physiology
  • Atlanto-Axial Joint* / surgery
  • Cadaver
  • Cervical Vertebrae / surgery
  • Feasibility Studies
  • Humans
  • Platybasia*
  • Spinal Fusion* / methods
  • Tomography, X-Ray Computed