A novel, highly sensitive electrochemical 1,4-dioxane sensor based on reduced graphene oxide-curcumin nanocomposite

RSC Adv. 2022 Jul 4;12(30):19375-19383. doi: 10.1039/d2ra01789j. eCollection 2022 Jun 29.

Abstract

1,4-Dioxane is a carcinogenic, non-biodegradable, organic water pollutant which is used as a solvent in various industries. It is also formed as an undesired by-product in the cosmetic and pharmaceutical industry. Given its carcinogenicity and ability to pollute, it is desirable to develop a sensitive and selective sensor to detect it in drinking water and other water bodies. Current works on this sensor are very few and involve complex metal oxide composite systems. A sensitive electrochemical sensor for 1,4-dioxane was developed by modifying a glassy carbon electrode (GCE) with a reduced graphene oxide-curcumin (rGO-CM) nanocomposite synthesized by a simple solution approach. The prepared rGO-CM was characterized by X-ray Diffraction (XRD), Fourier Transform Infrared (FTIR) Spectroscopy, Raman spectroscopy, UV-Vis spectroscopy, and Scanning Electron Microscopy (SEM). The rGO-CM/GCE sensor was employed for the detection of 1,4-dioxane in the range of 0.1-100 μM. Although, the detection range is narrower compared to reported literature, the sensitivity obtained for the proposed sensor is far superior. Moreover, the limit of detection (0.13 μM) is lower than the dioxane detection target defined by the World Health Organization (0.56 μM). The proposed rGO-CM/GCE also showed excellent stability and good recovery values in real sample (tap water and drinking water) analysis.