Long-Term Outcomes of the 150 mm Drug-Coated Balloon Cohort from the IN.PACT Global Study

Cardiovasc Intervent Radiol. 2022 Sep;45(9):1276-1287. doi: 10.1007/s00270-022-03214-y. Epub 2022 Jul 21.

Abstract

Purpose: Data on the long-term safety and effectiveness of drug-coated balloons (DCBs) for the treatment of long femoropopliteal atherosclerotic lesions in the real-world setting are rare. This study reports 3 year and 5 year outcomes of the pre-specified 150 mm balloon sub-cohort of the IN.PACT Global Study.

Methods: The IN.PACT Global Study was a prospective, multicentre, international, single-arm study evaluating the performance of the IN.PACT Admiral DCB in real-world patients with femoropopliteal atherosclerotic disease. This pre-specified 150 mm DCB cohort analysis comprised 107 participants (111 lesions) with all target lesions treated with at least one 150 mm DCB.

Results: Mean lesion length was 20.3 ± 9.2 cm; 18.0% had in-stent restenosis, 58.6% were totally occluded, and 17.1% were severely calcified. Through 60 months, the Kaplan-Meier estimate of freedom from clinically driven target lesion revascularization (CD-TLR) was 72.7% [95% confidence interval (CI):62.4%-80.5%]. The safety composite endpoint (freedom from device/procedure-related death through 30 days; freedom from target limb major amputation and clinically driven target vessel revascularization through 5 years) was 70.5%. The cumulative incidence of major amputation was 1.0% and all-cause mortality was 18.4% through 60 months. Freedom from CD-TLR rates in the provisional stented and non-stented subgroups through 36 months were 64.0% [95% CI: 46.1%-77.3%] and 81.9% [95% CI: 69.7%-89.6%] (log-rank p = 0.074), respectively.

Conclusions: The results demonstrate sustained long-term safety of the 150 mm IN.PACT Admiral DCB for long femoropopliteal atherosclerotic lesions in real-world patients. In particular, the results show that DCB angioplasty is an effective revascularization modality in long complex lesions.

Clinicaltrials: gov identifier: NCT01609296.

Level of evidence: Level 3, Cohort Study.

Keywords: Drug-coated balloon; Femoropopliteal; Long lesions; Peripheral artery disease; Target lesion revascularization.

MeSH terms

  • Angioplasty, Balloon*
  • Cardiovascular Agents*
  • Coated Materials, Biocompatible
  • Cohort Studies
  • Femoral Artery / diagnostic imaging
  • Humans
  • Paclitaxel
  • Peripheral Arterial Disease* / diagnostic imaging
  • Peripheral Arterial Disease* / therapy
  • Popliteal Artery / diagnostic imaging
  • Prospective Studies
  • Time Factors
  • Vascular Access Devices*
  • Vascular Patency

Substances

  • Cardiovascular Agents
  • Coated Materials, Biocompatible
  • Paclitaxel

Associated data

  • ClinicalTrials.gov/NCT01609296