The white blood cell differential. Evaluation of rapid impression scanning versus the routine manual count

Am J Clin Pathol. 1987 May;87(5):628-32. doi: 10.1093/ajcp/87.5.628.

Abstract

A quick, one and a half-minute qualitative microscopic scan was investigated as an alternative approach to the more labor-intensive 100-cell differential white blood cell count. The scanning results of 400 randomly selected hospital cases were compared with the on-line results of the 100-cell counts. Additionally, 50 cases selected to have a high percentage of abnormal results were each scanned and manually counted by four different readers. The results indicate that the scanning differential is equivalent to the 100-cell manual count in the detection of the presence of abnormal cell types such as immature granulocytes and blasts. Its ability to properly estimate the relative proportions of normal cells, especially lymphocytes, however, does not appear as reliable as the manual count. Most importantly, the analysis demonstrates that the scanning differential count exhibits a set of advantages and disadvantages that is complementary to those of the "three-part differential" technic provided by the newer generation automated hematologic analyzers. The authors therefore propose that these two procedures used in combination offer a suitable alternative to the manual 100-cell differential count.

Publication types

  • Comparative Study

MeSH terms

  • Basophils / cytology
  • Eosinophils / cytology
  • Humans
  • Leukocyte Count / methods*
  • Leukocyte Count / standards
  • Lymphocytes / cytology
  • Monocytes / cytology
  • Neutrophils / cytology