Primary Orbital Reconstruction with Selective Laser Melting (SLM) of Patient-Specific Implants (PSIs): An Overview of 96 Surgically Treated Patients

J Clin Med. 2022 Jun 11;11(12):3361. doi: 10.3390/jcm11123361.

Abstract

Contemporary advances in technology have allowed the transfer of knowledge from industrial laser melting systems to surgery; such an approach could increase the degree of accuracy in orbital restoration. The aim of this study was to examine the accuracy of selective laser melted PSIs (patient-specific implants) and navigation in primary orbital reconstruction. Ninety-six patients with orbital fractures were included in this study. Planned vs. achieved orbital volumes (a) and angles (b) were compared to the unaffected side (n = 96). The analysis included the overlay of post-treatment on planned images (iPlan 3.0.5, Brainlab®, Feldkirchen, Germany). The mean difference in orbital volume between the digitally planned orbit and the postoperative orbit was 29.16 cm3 (SD 3.54, presurgical) to 28.33 cm3 (SD 3.64, postsurgical, t = 5.00, df = 95.00; p < 0.001), resulting in a mean volume difference (planned vs. postop) of less than 1 cm3. A 3D analysis of the color mapping showed minor deviations compared to the mirrored unaffected side. The results suggested that primary reconstruction in complex orbital wall fractures can be routinely achieved with a high degree of accuracy by using selective laser melted orbital PSIs.

Keywords: 3D mesh; customized implant; intraoperative navigation; orbital reconstruction; orbital wall fracture; selective laser melting.

Grants and funding