Analytical performances of Exacto® HIV self-test in the Central African Republic

Pan Afr Med J. 2022 Mar 22:41:236. doi: 10.11604/pamj.2022.41.236.31220. eCollection 2022.

Abstract

We herein evaluated the analytical performances of the CE-IVD capillary blood Exacto® HIV self-test (Biosynex, Strasbourg, France) in the Central African Republic (CAR). A cross-sectional study was conducted on a representative national panel of 200 sera positive for HIV and 200 negative for HIV, randomly selected thorough the CAR for HIV seroprevalence surveillance survey, according to reference test. The Exacto® HIV self-test showed 99.5% (95% CI: 98.2-99.9) sensitivity and 100.0% (95% CI: 99.0-100.0) specificity. The Youden´s J index and Cohen´s Kappa coefficient were 0.995. At HIV-1 seroprevalence of 3.5% in the general adult population of the CAR, the positive and negative predictive values were 100% (95% CI: 99.0-100) and 99.9% (95% CI: 98.9-100), respectively. The results are within the limits required by the WHO (i.e. sensitivity ≥ 99.0% and specificity ≥ 98.0%), making Exacto® HIV self-test suitable for routine use in the CAR.

Keywords: Central African Republic; HIV; immunochromatography; rapid diagnostic test; self-testing.

MeSH terms

  • Adult
  • Central African Republic / epidemiology
  • Cross-Sectional Studies
  • HIV Infections* / diagnosis
  • HIV Infections* / epidemiology
  • Humans
  • Self-Testing
  • Sensitivity and Specificity
  • Seroepidemiologic Studies