Accuracy Evaluation of Collapsed Cone Convolution Superposition Algorithms for the Nasopharynx Interface in the Early Stage of Nasopharyngeal Carcinoma

Biomed Res Int. 2022 May 28:2022:5227609. doi: 10.1155/2022/5227609. eCollection 2022.

Abstract

This study combined the use of radiation dosimeteric measurements and a custom-made anthropomorphic phantom in order to evaluate the accuracy of therapeutic dose calculations at the nasopharyngeal air-tissue interface. The doses at the nasopharyngeal air-tissue interface obtained utilizing the Pinnacle and TomoTherapy TPS, which are based on collapsed cone convolution superposition (CCCS) algorithms, were evaluated and measured under single 10 × 10 cm2, 2 × 2 cm2, two parallel opposed 2 × 2 cm2 and clinical fields for early stage of nasopharyngeal carcinoma by using EBT3, GR-200F, and TLD 100. At the air-tissue interface under a 10 × 10 cm2 field, the TPS dose calculation values were in good agreement with the dosimeter measurement with all differences within 3.5%. When measured the single field 2 × 2 cm2, the differences between the average dose were measured at the distal interface for EBT3, GR-200F, and TLD-100 and the calculation values were -15.8%, -16.4%, and -4.9%, respectively. When using the clinical techniques such as IMRT, VMAT, and tomotherapy, the measurement results at the interface for all three techniques did not imply under dose. Small-field sizes will lead to dose overestimation at the nasopharyngeal air-tissue interface due to electronic disequilibrium when using CCCS algorithms. However, under clinical applications of multiangle irradiation, the dose errors caused by this effect were not significant.

MeSH terms

  • Algorithms
  • Humans
  • Nasopharyngeal Carcinoma / radiotherapy
  • Nasopharyngeal Neoplasms* / radiotherapy
  • Nasopharynx
  • Phantoms, Imaging
  • Radiometry / methods
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted* / methods