A convenient online desalination tube coupled with mass spectrometry for the direct detection of iodinated contrast media in untreated human spent hemodialysates

PLoS One. 2022 Jun 6;17(6):e0268751. doi: 10.1371/journal.pone.0268751. eCollection 2022.

Abstract

Background: Mass spectrometry (MS) analysis using direct infusion of biological fluids is often problematic due to high salts/buffers. Iodinated contrast media (ICM) are frequently used for diagnostic imaging purposes, sometimes inducing acute kidney injury (AKI) in patients with reduced kidney function. Therefore, detection of ICM in spent hemodialysates is important for AKI patients who require urgent continuous hemodiafiltration (CHDF) because it allows noninvasive assessment of the patient's treatment. In this study, we used a novel desalination tube before MS to inject the sample directly and detect ICM.

Methods: Firstly, spent hemodialysates of one patient were injected directly into the electrospray ionization (ESI) source equipped with a quadrupole time-of-flight mass spectrometer (Q-TOF MS) coupled to an online desalination tube for the detection of ICM and other metabolites. Thereafter, spent hemodialysates of two patients were injected directly into the ESI source equipped with a triple quadrupole mass spectrometer (TQ-MS) connected to that online desalination tube to confirm the detection of ICM.

Results: We detected iohexol (an ICM) from untreated spent hemodialysates of the patient-administered iohexol for computed tomography using Q-TOF MS. Using MRM profile analysis, we have confirmed the detection of ICM in the untreated spent hemodialysates of the patients administered for coronary angiography before starting CHDF. Using the desalination tube, we observed approximately 178 times higher signal intensity and 8 times improved signal-to-noise ratio for ioversol (an ICM) compared to data obtained without the desalination tube. This system was capable of tracking the changes of ioversol in spent hemodialysates of AKI patients by measuring spent hemodialysates.

Conclusion: The online desalination tube coupled with MS showed the capability of detecting iohexol and ioversol in spent hemodialysates without additional sample preparation or chromatographic separation. This approach also demonstrated the capacity to monitor the ioversol changes in patients' spent hemodialysates.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Acute Kidney Injury* / chemically induced
  • Acute Kidney Injury* / diagnosis
  • Contrast Media / adverse effects
  • Hemodialysis Solutions
  • Humans
  • Iodine Compounds*
  • Iohexol
  • Mass Spectrometry
  • Spectrometry, Mass, Electrospray Ionization

Substances

  • Contrast Media
  • Hemodialysis Solutions
  • Iodine Compounds
  • Iohexol

Grants and funding

This research work was supported by MEXT Project for promoting public utilization of advanced research infrastructure (Imaging Platform) (Grant number JPMXS0410300220), Japan.