Bioresorbable Nanostructured Chemical Sensor for Monitoring of pH Level In Vivo

Adv Sci (Weinh). 2022 Aug;9(22):e2202062. doi: 10.1002/advs.202202062. Epub 2022 May 26.

Abstract

Here, the authors report on the manufacturing and in vivo assessment of a bioresorbable nanostructured pH sensor. The sensor consists of a micrometer-thick porous silica membrane conformably coated layer-by-layer with a nanometer-thick multilayer stack of two polyelectrolytes labeled with a pH-insensitive fluorophore. The sensor fluorescence changes linearly with the pH value in the range 4 to 7.5 upon swelling/shrinking of the polymer multilayer and enables performing real-time measurements of the pH level with high stability, reproducibility, and accuracy, over 100 h of continuous operation. In vivo studies carried out implanting the sensor in the subcutis on the back of mice confirm real-time monitoring of the local pH level through skin. Full degradation of the pH sensor occurs in one week from implant in the animal model, and its biocompatibility after 2 months is confirmed by histological and fluorescence analyses. The proposed approach can be extended to the detection of other (bio)markers in vivo by engineering the functionality of one (at least) of the polyelectrolytes with suitable receptors, thus paving the way to implantable bioresorbable chemical sensors.

Keywords: bioresorbable chemical sensor; fluorescence sensor; implanted chemical sensor; in vivo pH sensor; layer-by-layer (LbL); porous silicon; wireless sensor.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Absorbable Implants*
  • Animals
  • Hydrogen-Ion Concentration
  • Mice
  • Nanostructures*
  • Polyelectrolytes
  • Reproducibility of Results

Substances

  • Polyelectrolytes