Gaseous oxidized mercury (Hg2+) monitoring is one of the largest challenges in the mercury research field, where existing methods cannot simultaneously satisfy the measurement requirements of both accuracy and time precision, especially in high-particulate environments. Here, we verified that dual-stage cation exchange membrane (CEM) sampler is incapable of gaseous elemental mercury (Hg0) uptake even if particulate matter is trapped on CEM, whereas the Hg2+ capture efficiency of the sampler is more than 90%. We then developed a Cation Exchange Membrane-Coupled Speciated Atmospheric Mercury Monitoring System (CSAMS) by coupling the dual-stage CEM sampler with the commercial Tekran 2537/1130/1135 system and configuring a new sampling and analysis procedure, so as to improve the monitoring accuracy of Hg2+ and ensure the simultaneous measurement of Hg0, Hg2+, and Hgp in 2 h time resolution. We deployed the CSAMS in urban Beijing in September 2021 and observed an unprecedented elevated Hg2+ during the daytime with an average amplitude of 510 pg m-3. Using a zero-dimensional box model, the elevated Hg2+ production rate was attributed to high atmospheric oxidant concentrations, Hg0 heterogeneous and interfacial oxidation processes on the surface of atmospheric particles, or potential unknown oxidants.
Keywords: cation exchange membrane; gaseous oxidized mercury; online monitoring; photochemical oxidation; speciated atmospheric mercury.