Trifecta versus Perimount Magna Ease aortic valves: Failure mechanisms

Asian Cardiovasc Thorac Ann. 2022 Sep;30(7):797-806. doi: 10.1177/02184923221100994. Epub 2022 May 22.

Abstract

Background: There are increasing reports of early externally mounted pericardial Trifecta bioprosthesis failure. We compared the hemodynamic performance of Trifecta and Carpentier-Edwards Perimount Magna Ease valves to determine the failure mechanism.

Methods: We retrospectively included 270 consecutive patients (age: 73.4 ± 8.2 years; 57.5% male; mean follow-up: 48.0 ± 20.3 months) who underwent aortic valve replacement from 2014 to 2021 at a single center and compared the Trifecta (N = 137) and Carpentier-Edwards Perimount Magna Ease valve (N = 133) patients.

Results: The prosthetic valve major aortic regurgitation incidence was higher for the Trifecta than that for the Carpentier-Edwards Perimount Magna Ease valve (6.3% vs. 0%, P < 0.009). Among the Trifecta failures, 33% developed structural valve deterioration, but all requiring redo aortic valve replacement developed major prosthetic valve aortic regurgitation. Freedom at 5 years from redo aortic valve replacement due to structural valve deterioration was significantly lower for Trifecta (89.4% vs. 100%, P = 0.003). The reoperation hazards were determined for Trifecta (vs. Carpentier-Edwards Perimount Magna Ease): 11.6 (1.47-90.9; P = 0.02), prosthetic valve aortic regurgitation: 2.38 (1.70-3.32; P < 0.01), structural valve deterioration: 20.82 (4.08-106.2; P < 0.01), 5-year mean transprosthetic pressure gradient: 1.14 per 1-point increase (1.03-1.24; P = 0.007), and urgent surgery: 10.1 (2.59-39.0; P = 0.001). The Cox regression analysis identified that prosthetic valve aortic regurgitation solely contributed to redo aortic valve replacement (hazard ratio: 2.38; confidence intervals: 1.70-3.32).

Conclusions: Significantly, more early failures occurred with the Trifecta valve than the Carpentier-Edwards Perimount Magna Ease valve but the Trifecta showed reasonable mean transprosthetic pressure gradient over time. Prosthetic valve aortic regurgitation and calcific structural valve deterioration synergistically contributed to Trifecta valve failure alternatively.

Keywords: Magna ease; Trifecta; aortic prosthetic valve regurgitation; aortic valve replacement; reoperation; structural valve deterioration.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / surgery
  • Aortic Valve Insufficiency* / diagnostic imaging
  • Aortic Valve Insufficiency* / etiology
  • Aortic Valve Insufficiency* / surgery
  • Bioprosthesis*
  • Female
  • Heart Valve Prosthesis Implantation* / adverse effects
  • Heart Valve Prosthesis*
  • Hemodynamics
  • Humans
  • Male
  • Prosthesis Design
  • Retrospective Studies
  • Treatment Outcome