Spray Pattern and Plume Geometry Testing and Methodology: An IPAC-RS Working Group Overview

AAPS PharmSciTech. 2022 May 18;23(5):145. doi: 10.1208/s12249-022-02278-w.

Abstract

Plume characterization for orally inhaled and nasal drug products (OINDP) provides valuable information during OINDP development. Spray pattern and plume geometry techniques, methods, and technology have evolved over the past 20 years since the publication of the original 1998 FDA MDI DPI draft guidance. The International Pharmaceutical Aerosol Consortium on Regulation and Science (IPAC-RS) discusses the historical context and background to plume geometry and spray pattern characterization studies; provides an analysis of the current regulatory context; addresses results from its industry surveys on application and value of such testing; and presents case studies and best practices-seeking to provide insights to regulatory bodies and other stakeholders. Assessment and consideration of published studies and industry experience note the value of plume geometry and spray pattern in development, and that further data is needed regarding their use in assessing formulation characteristics. Continued dialogue between industry and regulatory bodies is needed to establish the optimum use of these techniques.

Keywords: inhalation; nasal; plume geometry; spray pattern.

Publication types

  • Review

MeSH terms

  • Administration, Inhalation
  • Administration, Intranasal*
  • Aerosols
  • Particle Size

Substances

  • Aerosols