The Use of a Commercially Available Endovascular Filter Catheter (Capturex®) for Thoracic Endovascular Aortic Repair (TEVAR) in a Patient with a Coral Reef Aorta

Cardiovasc Intervent Radiol. 2022 Jun;45(6):879-883. doi: 10.1007/s00270-022-03147-6. Epub 2022 Apr 26.

Abstract

Due to the risk of mobilizing plaque fragments, transfemoral TEVAR is a potentially dangerous procedure in patients with a coral reef aorta. We describe a practical method for transfemoral TEVAR in a patient with a degenerative thoracic aneurysm and a coral reef aorta. After placing a filter catheter in the abdominal aorta via a contralateral percutaneous femoral access, a working channel through the distal thoracic aorta was created with a balloon-expandable stent graft in the coral reef segment. Thereafter, transfemoral TEVAR could be performed successfully, without any complications. The additional use of a percutaneously placed filter catheter potentially allows reduction of peripheral embolism and hence may prevent patients from more invasive treatment.

Keywords: Coral reef aorta; Filter catheter; TEVAR; Thoracic aortic aneurysm.

MeSH terms

  • Aorta
  • Aorta, Thoracic / diagnostic imaging
  • Aorta, Thoracic / surgery
  • Aortic Aneurysm, Thoracic* / complications
  • Aortic Aneurysm, Thoracic* / diagnostic imaging
  • Aortic Aneurysm, Thoracic* / surgery
  • Blood Vessel Prosthesis
  • Blood Vessel Prosthesis Implantation* / adverse effects
  • Catheters
  • Coral Reefs
  • Endovascular Procedures* / methods
  • Humans
  • Retrospective Studies
  • Stents
  • Treatment Outcome