Technical note: Characterization and practical applications of a novel plastic scintillator for online dosimetry for an ultrahigh dose rate (FLASH)

Med Phys. 2022 Jul;49(7):4682-4692. doi: 10.1002/mp.15671. Epub 2022 May 4.

Abstract

Purpose: Although flash radiation therapy (FLASH-RT) is a promising novel technique that has the potential to achieve a better therapeutic ratio between tumor control and normal tissue complications, the ultrahigh pulsed dose rates (UHPDR) mean that experimental dosimetry is very challenging. There is a need for real-time dosimeters in the development and implementation of FLASH-RT. In this work, we characterize a novel plastic scintillator capable of temporal resolution short enough (2.5 ms) to resolve individual pulses.

Methods: We characterized a novel plastic dosimeter for use in a linac converter to deliver 16-MeV electrons at 100-Gy/s UHPDR average dose rates. The linearity and reproducibility were established by comparing relative measurements with a pinpoint ionization chamber placed at 10-cm water-equivalent depth where the electrometer is not saturated by the high dose per pulse. The accuracy was established by comparing the plastic scintillator dose measurements with EBT-XD Gafchromic radiochromic films, the current reference dosimeter for UHPDR. Finally, the plastic scintillator was compared against EBT-XD films for online dosimetry of two in vitro experiments performed at UHPDR.

Results: Relative ion chamber measurements were linear with plastic scintillator response within ≤1% over 4-20 Gy and pulse frequencies (18-180 Hz). When characterized under reference conditions with NIST-traceability, the plastic scintillator maintained its dose response under UHPDR conditions and agreed with EBT-XD film dose measurements within 4% under reference conditions and 6% for experimental online dosimetry.

Conclusion: The plastic scintillator shows a linear and reproducible response and is able to accurately measure the radiation absorbed dose delivered by 16-MeV electrons at UHPDR. The dose is measured accurately in real time with a greater level of precision than that achieved with a radiochromic film.

Keywords: FLASH-RT; scintillator detectors; small animal RT.

MeSH terms

  • Electrons
  • Film Dosimetry / methods
  • Plastics*
  • Radiation Dosage
  • Radiometry*
  • Reproducibility of Results