Investigation of the influence of blade configuration on the hemodynamic performance and blood damage of the centrifugal blood pump

Artif Organs. 2022 Sep;46(9):1817-1832. doi: 10.1111/aor.14265. Epub 2022 Apr 25.

Abstract

Purpose: The design and optimization of centrifugal blood pumps are crucial for improved extracorporeal membrane oxygenation system performance. Secondary flow passages are common in centrifugal blood pumps, allowing for a high volume of unstable flow. Traditional design theory offers minimal guidance on the design and optimization of centrifugal blood pumps, so it's critical to understand how design parameter variables affect hydraulic performance and hemocompatibility.

Methods: Computational fluid dynamics (CFD) was employed to investigate the effects of blade number, blade wrap angle, blade thickness, and splitters on pressure head, hemolysis, and platelet activation state. Eulerian and Lagrangian features were used to analyze the flow fields and hemocompatibility metrics such as scalar shear stress, velocity distribution, and their correlation.

Results: The equalization of frictional and flow losses allow impellers with more blades and smaller wrap angles to have higher pressure heads, whereas the trade-off between the volume of high scalar shear stress and exposure time allows impellers with fewer blades and larger blade wrap angles to have a lower HI; there are configurations that increase the possibility of platelet activation for both number of blades and wrap angles. The hydraulic performance and hemocompatibility of centrifugal blood pumps are not affected by blade thickness. Compared to the main blades, splitters can improve the blood compatibility of a centrifugal blood pump with a small reduction in pressure head, but there is a trade-off between the length and location of the splitter that suppresses flow losses while reducing the velocity gradient. According to correlation analysis, pressure head, HI, and the volume of high shear stress were all substantially connected, and exposure time had a significant impact on HI. The platelet activation state was influenced by the average scalar shear stress and the volume of low velocity.

Conclusion: The findings reveal the impact of design variables on the performance of centrifugal blood pumps with secondary flow passages, as well as the relationship between hemocompatibility, hydraulic performance, and flow characteristics, and are useful for the design and optimization of this type of blood pump, as well as the prediction of clinical complications.

Keywords: blade configuration; centrifugal blood pump; hemolysis; platelet activation state; pressure head.

MeSH terms

  • Equipment Design
  • Heart-Assist Devices* / adverse effects
  • Hemodynamics
  • Hemolysis / physiology
  • Humans
  • Hydrodynamics