Clinical Performance of the cobas Liat SARS-CoV-2 & Influenza A/B Assay in Nasal Samples

Mol Diagn Ther. 2022 May;26(3):323-331. doi: 10.1007/s40291-022-00580-8. Epub 2022 Apr 7.

Abstract

Background and objective: Point-of-care type molecular diagnostic tests have been used for detecting SARS-CoV-2, although their clinical utility with nasal samples has yet to be established. This study evaluated the clinical performance of the cobas Liat SARS-CoV-2 & Influenza A/B (Liat) assay in nasal samples.

Methods: Nasal and nasopharyngeal samples were collected and were tested using the Liat, the cobas 6800 system and the cobas SARS-CoV-2 & Influenza A/B (cobas), and a method developed by National Institute of Infectious Diseases, Japan (NIID).

Results: A total of 814 nasal samples were collected. The Liat assay was positive for SARS-CoV-2 in 113 (13.9%). The total, positive, and negative concordance rate between the Liat and cobas/NIID assays were 99.3%/98.4%, 99.1%/100%, and 99.3%/98.2%, respectively. Five samples were positive only using the Liat assay. Their Ct values ranged from 31.9 to 37.2. The Ct values of the Liat assay were significantly lower (p < 0.001) but were correlated (p < 0.001) with those of other molecular assays. In the participants who tested positive for SARS-CoV-2 on the Liat assay using nasopharyngeal samples, 88.2% of their nasal samples also tested positive using the Liat assay.

Conclusion: The Liat assay showed high concordance with other molecular assays in nasal samples. Some discordance occurred in samples with Ct values > 30 on the Liat assay.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • COVID-19* / diagnosis
  • Humans
  • Influenza, Human* / diagnosis
  • Nasopharynx
  • SARS-CoV-2
  • Sensitivity and Specificity