Prepolarized MRI of hard tissues and solid-state matter

NMR Biomed. 2022 Aug;35(8):e4737. doi: 10.1002/nbm.4737. Epub 2022 Apr 20.

Abstract

Prepolarized MRI (PMRI) is a long-established technique conceived to counteract the loss in signal-to-noise ratio (SNR) inherent to low-field MRI systems. When it comes to hard biological tissues and solid-state matter, PMRI is severely restricted by their ultra-short characteristic relaxation times. Here we demonstrate that efficient hard-tissue prepolarization is within reach with a special-purpose 0.26 T scanner designed for ex vivo dental MRI and equipped with suitable high-power electronics. We have characterized the performance of a 0.5 T prepolarizer module, which can be switched on and off in 200 μs. To this end, we have used resin, dental and bone samples, all with T1 times of the order of 20 ms at our field strength. The measured SNR enhancement is in good agreement with a simple theoretical model, and deviations in extreme regimes can be attributed to mechanical vibrations due to the magnetic interaction between the prepolarization and main magnets.

Keywords: MRI; hard tissues; low field; prepolarization; solid state.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Magnetic Resonance Imaging* / methods
  • Magnetics*
  • Magnets
  • Models, Theoretical
  • Signal-To-Noise Ratio