Analysis of the accuracy of automatic electrocardiogram interpretation in ST-segment elevation myocardial infarction

Clin Exp Emerg Med. 2022 Mar;9(1):18-23. doi: 10.15441/ceem.21.163. Epub 2022 Mar 31.

Abstract

Objective: This study aimed to analyze the association between the culprit artery and the diagnostic accuracy of automatic electrocardiogram (ECG) interpretation in patients with ST-segment elevation myocardial infarction (STEMI).

Methods: This single-centered, retrospective cohort study included adult patients with STEMI who visited the emergency department between January 2017 and December 2020. The primary endpoint was the association between the culprit artery occlusion and the misinterpretation of ECG, evaluated by the chi-square test or Fisher exact test.

Results: The rate of misinterpretation of the automated ECG for patients with STEMI was 26.5% (31/117 patients). There was no significant correlation between the ST segment change in the four involved leads (anteroseptal, lateral, inferior, and aVR) and the misinterpretation of ECG (all P > 0.05). Single culprit artery occlusion significantly affected the misinterpretation of ECG compared with multiple culprit artery occlusion (single vs. multiple, 27/86 [31.3%] vs. 4/31 [12.9%], P = 0.045). There was no association between culprit artery and the misinterpretation of ECG (P = 0.132).

Conclusion: Single culprit artery occlusion might increase misinterpretation of ECG compared with multiple culprit artery occlusions in the automatic interpretation of STEMI.

Keywords: Coronary vessels; Diagnosis; Electrocardiography; ST elevation myocardial infarction.