Reliability and benefits of single-energy projection-based metallic artifact reduction (SEMAR) in the different orthopedic hardware for the hip

Skeletal Radiol. 2022 Sep;51(9):1853-1863. doi: 10.1007/s00256-022-04040-6. Epub 2022 Mar 26.

Abstract

Objective: To evaluate the performance and reliability of the single-energy metal artifact reduction (SEMAR) algorithm in patients with different orthopedic hardware at the hips.

Materials and methods: A total of 153 patients with hip instrumentation who had undergone CT with adaptive iterative dose reduction (AIDR) 3D and SEMAR algorithms between February 2015 and October 2019 were included retrospectively. Patients were divided into 5 groups by the hardware type. Two readers with 21 and 13 years of experience blindly reviewed all image sets and graded the extent of artifacts and imaging quality using 5-point scales. To evaluate reliability, the mean densities and image noises were measured at the urinary bladder, veins, and fat in images with artifacts and the reference images.

Results: No significant differences were found in the mean densities of the urinary bladder, veins, and fat between the SEMAR images with artifacts (7.57 ± 9.49, 40.29 ± 23.07, 86.78 ± 38.34) and the reference images (7.77 ± 6.2, 40.27 ± 8.66, 89.10 ± 20.70) (P = .860, .994, .392). Image noises of the urinary bladder in the SEMAR images with artifacts (14.25 ± 4.50) and the SEMAR reference images (9.69 ± 1.29) were significantly higher than those in the AIDR 3D reference images (9.11 ± 1.12) (P < .001; P < .001). All AIDR 3D images were non-diagnostic (overall quality ≤ 3) and less than a quarter of the SEMAR images were non-diagnostic (16.7-23.7%), mainly in patients with prostheses [reader 1: 91.7% (22/24); reader 2: 92.6% (25/27)].

Conclusion: The SEMAR algorithm significantly reduces metal artifacts in CT images, more in patients with internal fixations than in patients with prostheses, and provides reliable attenuation of soft tissues.

Keywords: Computed tomography (CT); Hip; Metal artifact; Metal artifact reduction; Orthopedic hardware.

MeSH terms

  • Algorithms
  • Artifacts*
  • Humans
  • Image Processing, Computer-Assisted* / methods
  • Metals
  • Reproducibility of Results
  • Retrospective Studies
  • Tomography, X-Ray Computed / methods

Substances

  • Metals