Development, Monte Carlo simulations and experimental evaluation of a 3D range-modulator for a complex target in scanned proton therapy

Biomed Phys Eng Express. 2022 Mar 11;8(3). doi: 10.1088/2057-1976/ac5937.

Abstract

The purpose of this work was to develop and manufacture a 3D range-modulator (3D RM) for a complex target contour for scanned proton therapy. The 3D RM is considered to be a viable technique for the very fast dose application in patient-specific tumors with only one fixed energy. The RM was developed based on a tumor from a patient CT and manufactured with high-quality 3D printing techniques with both polymer resin and aluminum. Monte Carlo simulations were utilized to investigate its modulating properties and the resulting dose distribution. Additionally, the simulation results were validated with measurements at the Marburg Ion-Beam Therapy Centre. For this purpose, a previously developed water phantom was used to conduct fast, automated high-resolution dose measurements. The results show a very good agreement between simulations and measurements and indicate that highly homogeneous dose distributions are possible. The delivered dose is conformed to the distal as well as to the proximal edge of the target. The 3D range-modulator concept combines a high degree of dose homogeneity and conformity, comparable to standard IMPT with very short irradiation times, promising clinically applicable dose distributions for lung and/or FLASH treatment, comparable and competitive to those from conventional irradiation techniques.

Keywords: 3D printing; 3D range-modulator; FLUKA; monte carlo simulation; particle therapy; proton therapy.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Computer Simulation
  • Humans
  • Monte Carlo Method
  • Phantoms, Imaging
  • Proton Therapy*
  • Water

Substances

  • Water