Design of an Effective Prosthetic Hand System for Adaptive Grasping with the Control of Myoelectric Pattern Recognition Approach

Micromachines (Basel). 2022 Jan 29;13(2):219. doi: 10.3390/mi13020219.

Abstract

In this paper, we develop a prosthetic bionic hand system to realize adaptive gripping with two closed-loop control loops by using a linear discriminant analysis algorithm (LDA). The prosthetic hand contains five fingers and each finger is driven by a linear servo motor. When grasping objects, four fingers except the thumb would adjust automatically and bend with an appropriate gesture, while the thumb is stretched and bent by the linear servo motor. Since the change of the surface electromechanical signal (sEMG) occurs before human movement, the recognition of sEMG signal with LDA algorithm can help to obtain people's action intention in advance, and then timely send control instructions to assist people to grasp. For activity intention recognition, we extract three features, Variance (VAR), Root Mean Square (RMS) and Minimum (MIN) for recognition. As the results show, it can achieve an average accuracy of 96.59%. This helps our system perform well for disabilities to grasp objects of different sizes and shapes adaptively. Finally, a test of the people with disabilities grasping 15 objects of different sizes and shapes was carried out and achieved good experimental results.

Keywords: LDA; motion recognition; prosthetic bionic hand; sEMG signal.