Comparison of peak skin dose and dose map obtained with real-time software and radiochromic films in patients undergoing abdominopelvic embolization

Diagn Interv Imaging. 2022 Jul-Aug;103(7-8):338-344. doi: 10.1016/j.diii.2022.01.010. Epub 2022 Feb 17.

Abstract

Purpose: The purpose of this study was to compare peak skin dose (PSD) and dose map calculated by Dose-Tracking-System® (DTS) software and measured with radiochromic films in patients undergoing abdominopelvic embolization.

Material and methods: The PSD measured by radiochromic films (PSDFilm) or calculated by DTS software (PSDDTS) were compared in patients who underwent abdominopelvic embolization between September 2020 and April 2021. Concordance between PSDFilm and PSDDTS was computed using the Lin's concordance correlation coefficient and the clinical concordance using Bland Altman analysis. PSD values were compared using the paired Mann-Whitney-Wilcoxon test.

Results: A total of 40 patients were included. There were 32 men and 8 women with a mean age of 73.0 ± 14.6 (SD) years (age range: 30-92 years). Median PSDFilm was 756 mGy (IQR: 390; 1094) and median PSDDTS was 768 mGy (IQR: 421; 1076), resulting in a median difference of -5% (IQR: -10%; 0%) between PSDFilm and PSDDTS (P = 0.024). The concordance correlation between PSDFilm and PSDDTS was substantial in patients (0.986; 95% CI: 0.977-0.992). Bland Altman analysis showed that PSDDTS was underestimated compared to PSDFilm by -36 mGy (95% CI: -68--4). Visually similar dose maps were found with DTS and radiochromic films.

Conclusion: Real-time DTS software allows computing PSD with high accuracy and generating adequate dose map. PSDDTS was slightly underestimated compared to PSDFilm requiring the use of a correction factor for the PSDDTS to avoid lack of follow-up for some patients.

Keywords: Embolization; Film dosimetry; Fluoroscopy; Interventional radiology; Radiochromic films.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Embolization, Therapeutic* / methods
  • Female
  • Film Dosimetry* / methods
  • Humans
  • Male
  • Middle Aged
  • Radiation Dosage
  • Skin
  • Software