Objectives: This study was conducted to evaluate the effect of dose reduction on the performance of a deep learning (DL)-based computer-aided diagnosis (CAD) system regarding pulmonary nodule detection in a virtual screening scenario.
Methods: Sixty-eight anthropomorphic chest phantoms were equipped with 329 nodules (150 ground glass, 179 solid) with four sizes (5 mm, 8 mm, 10 mm, 12 mm) and scanned with nine tube voltage/current combinations. The examinations were analyzed by a commercially available DL-based CAD system. The results were compared by a comparison of proportions. Logistic regression was performed to evaluate the impact of tube voltage, tube current, nodule size, nodule density, and nodule location.
Results: The combination with the lowest effective dose (E) and unimpaired detection rate was 80 kV/50 mAs (sensitivity: 97.9%, mean false-positive rate (FPR): 1.9, mean CTDIvol: 1.2 ± 0.4 mGy, mean E: 0.66 mSv). Logistic regression revealed that tube voltage and current had the greatest impact on the detection rate, while nodule size and density had no significant influence.
Conclusions: The optimal tube voltage/current combination proposed in this study (80 kV/50 mAs) is comparable to the proposed combinations in similar studies, which mostly dealt with conventional CAD software. Modification of tube voltage and tube current has a significant impact on the performance of DL-based CAD software in pulmonary nodule detection regardless of their size and composition.
Key points: • Modification of tube voltage and tube current has a significant impact on the performance of deep learning-based CAD software. • Nodule size and composition have no significant impact on the software's performance. • The optimal tube voltage/current combination for the examined software is 80 kV/50 mAs.
Keywords: Artificial intelligence; Computer-assisted diagnosis; Deep learning; Lung neoplasms; Radiographic phantoms.