Satisfied quantitative value can be acquired by short-time bone SPECT/CT using a whole-body cadmium-zinc-telluride gamma camera

Sci Rep. 2021 Dec 21;11(1):24320. doi: 10.1038/s41598-021-03853-0.

Abstract

The aim of this study was to evaluate the quantitative values of short-time scan (STS) of metastatic lesions compared with a standard scan (SS) when acquired by whole-body bone SPECT/CT with cadmium-zinc-telluride (CZT) detectors. We retrospectively reviewed 13 patients with bone metastases from prostate cancer, who underwent SPECT/CT performed on whole-body CZT gamma cameras. STSs were obtained using 75, 50, 25, 10, and 5% of the list-mode data for SS, respectively. Regions of interest (ROIs) were set on the increased uptake areas diagnosed as metastases. Intraclass correlation coefficients (ICCs) of standardized uptake values (SUVs) for the ROIs were calculated between the SS and each STS, and ICC ≥ 0.8 was set as a perfect correlation. Moreover, the repeatability coefficient (RC) was calculated, and RC ≤ 20% was defined as acceptable. A total of 152 metastatic lesions were included in the analysis. The ICCs between the SS vs. 75%-STS, 50%-STS, 25%-STS, 10%-STS, and 5%-STS were 0.999, 0.997, 0.994, 0.983, and 0.955, respectively. The RCs of the SS vs. 75%-STS, 50%-STS, 25%-STS, 10%-STS, and 5%-STS were 7.9, 12.4, 19.8, 30.8, and 41.3%, respectively. When evaluating the quality of CZT bone SPECT/CT acquired by a standard protocol, 25%-STS may provide adequate quantitative values.