Three-dimensional reduced field-of-view imaging (3D-rFOVI)

Magn Reson Med. 2022 May;87(5):2372-2379. doi: 10.1002/mrm.29121. Epub 2021 Dec 11.

Abstract

Purpose: This study aimed at developing a 3D reduced field-of-view imaging (3D-rFOVI) technique using a 2D radiofrequency (RF) pulse, and demonstrating its ability to achieve isotropic high spatial resolution and reduced image distortion in echo planar imaging (EPI).

Methods: The proposed 3D-rFOVI technique takes advantage of a 2D RF pulse to excite a slab along the conventional slice-selection direction (i.e., z-direction) while limiting the spatial extent along the phase-encoded direction (i.e., y-direction) within the slab. The slab is phase-encoded in both through-slab and in-slab phase-encoded directions. The 3D-rFOVI technique was implemented at 3T in gradient-echo and spin-echo EPI pulse sequences for functional MRI (fMRI) and diffusion-weighted imaging (DWI), respectively. 3D-rFOVI experiments were performed on a phantom and human brain to illustrate image distortion reduction, as well as isotropic high spatial resolution, in comparison with 3D full-FOV imaging.

Results: In both the phantom and the human brain, image voxel dislocation was substantially reduced by 3D-rFOVI when compared with full-FOV imaging. In the fMRI experiment with visual stimulation, 3D isotropic spatial resolution of (2 × 2 × 2 mm3 ) was achieved with an adequate signal-to-noise ratio (81.5) and blood oxygen level-dependent (BOLD) contrast (2.5%). In the DWI experiment, diffusion-weighted brain images with an isotropic resolution of (1 × 1 × 1 mm3 ) was obtained without appreciable image distortion.

Conclusion: This study indicates that 3D-rFOVI is a viable approach to 3D neuroimaging over a zoomed region.

Keywords: 2D RF pulse; 3D imaging; DWI; EPI; fMRI; isotropic resolution; reduced FOV.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Diffusion Magnetic Resonance Imaging* / methods
  • Echo-Planar Imaging* / methods
  • Humans
  • Imaging, Three-Dimensional / methods
  • Phantoms, Imaging
  • Signal-To-Noise Ratio