Partial nephrectomy through retroperitoneal approach with a new surgical robot system, KD-SR-01

Int J Med Robot. 2022 Apr;18(2):e2352. doi: 10.1002/rcs.2352. Epub 2021 Nov 20.

Abstract

Background: To present our experiences with partial nephrectomy (PN) through retroperitoneal approach (RP) with the Kangduo robotic system.

Methods: From December 2020 to February 2021, the perioperative data of 11 patients underwent PN through RP with the Kangduo robotic system were collected prospectively.

Results: For the R.E.N.A.L. nephrometry score, 72.7% of patients had a low score (4-6) and 27.3% of patients had a medium score (7-9). Seven tumours were posterior (P), four tumours were on the midline (X). All procedures were completed successfully. The median warm ischemia time was 18.5 (IQR, 13.7-21.0) min. None of the patients had positive surgical margins at definitive histology (all pT1a). No high-grade perioperative complications or device-related adverse events occurred. At a mean follow-up of 8 ± 0.8 months, no complications occurred in all patients.

Conclusions: RPPN using the novel Kangduo robotic system is a safe and effective option for managing posterior and lateral renal tumours with R.E.N.A.L. nephrometry scores ≤9.

Keywords: KD-SR-01; partial nephrectomy; retroperitoneal approach; surgical robot.

MeSH terms

  • Humans
  • Kidney Neoplasms* / pathology
  • Kidney Neoplasms* / surgery
  • Nephrectomy / methods
  • Operative Time
  • Retrospective Studies
  • Robotic Surgical Procedures* / methods
  • Robotics*
  • Treatment Outcome