Cow-hitch fixation in fracture hemiarthroplasty

JSES Int. 2021 Sep 11;5(6):1027-1033. doi: 10.1016/j.jseint.2021.07.011. eCollection 2021 Nov.

Abstract

Background: The treatment of complex proximal humerus fractures with hemiarthroplasty is associated with a high failure rate due to secondary displacement of the tuberosities. It was the aim of this in-vitro study to compare the mechanical stability of tuberosity reattachment obtained with the so-called "Cow-Hitch" (CH) cerclage compared with conventional tuberosity reattachment.

Methods: A 4-part proximal humerus fracture was created in 10 fresh-frozen, human cadaveric shoulders. The greater and lesser tuberosity were reattached to the hemiarthroplasty stem with in total 4 CH Cerclages in the Cow-Hitch group. The conventional technique-recommended for the tested implant-was used in the control group using 6 sutures. A total of 5000 loading cycles with forces of 350N were applied, while motion (in mm) of the tuberosities was recorded in 3 directions (anteroposterior = AP, mediolateral = ML, inferosuperior = IS) with a telecentric camera.

Results: After 5000 loading cycles, the CH group showed less fragment displacement (AP: 2.3 ± 2.3 mm, ML: 1.8 ± 0.9 mm, IS: 1.3 ± 0.5 mm) than the conventional group (AP: 9.8 ± 12.3 mm, ML: 5.5 ± 5.6 mm, IS: 4.5 ± 4.7 mm). The differences were not statistically significant (AP: P = .241; ML: P = .159; IS: P = .216). The lesser tuberosity fragment displacement in the CH group after 5000 cycles was less in the AP (2.3 ± 3.3 vs. 4.0 ± 2.8, P = .359) and IS (1.9 ± 1.2 vs. 3.1 ± 1.8; P = .189) directions but higher in the ML direction (7.2 ± 5.7 vs 6.3 ± 3.6, P = .963).

Conclusions: In-vitro, "Cow-Hitch" cerclage results in mean greater tuberosity displacements of 2 mm and reliably prevents displacements greater than 5 mm. In contrast, the conventional fixation technique yields unreliable, variable stability with low to complete displacement upon cyclical loading.

Keywords: Cadaver study; Cow Hitch Cerclage; Greater tuberosity reattachment; Hemiarthroplasty; Lesser tuberosity reattachment; Proximal humerus fracture.