Accuracy of Flash Glucose Monitoring During Postprandial Rest and Different Walking Conditions in Overweight or Obese Young Adults

Front Physiol. 2021 Oct 15:12:732751. doi: 10.3389/fphys.2021.732751. eCollection 2021.

Abstract

Aims: To investigate the accuracy of FreeStyle LibreTM flash glucose monitoring (FGM) relevant to plasma glucose (PG) measurements during postprandial rest and different walking conditions in overweight/obese young adults. Methods: Data of 40 overweight/obese participants from two randomized crossover studies were pooled into four trials: (1) sitting (SIT, n = 40); (2) walking continuously for 30 min initiated 20 min before individual postprandial glucose peak (PPGP) (20iP + CONT, n = 40); (3) walking continuously for 30 min initiated at PPGP (iP + CONT, n = 20); and (4) accumulated walking for 30 min initiated 20 min before PPGP (20iP + ACCU, n = 20). Paired FGM and PG were measured 4 h following breakfast. Results: The overall mean absolute relative difference (MARD) between PG and FGM readings was 16.4 ± 8.6% for SIT, 16.2 ± 4.7% for 20iP + CONT, 16.7 ± 12.2% for iP + CONT, and 19.1 ± 6.8% for 20iP + ACCU. The Bland-Altman analysis showed a bias of -1.03 mmol⋅L-1 in SIT, -0.89 mmol⋅L-1 in 20iP + CONT, -0.82 mmol⋅L-1 in iP + CONT, and -1.23 mmol⋅L-1 in 20iP + ACCU. The Clarke error grid analysis showed that 99.6-100% of the values in all trials fell within zones A and B. Conclusion: Although FGM readings underestimated PG, the FGM accuracy was overall clinically acceptable during postprandial rest and walking in overweight/obese young adults.

Keywords: continuous glucose monitoring; exercise; obesity; postprandial glycemia; sensor accuracy.