A sensitive LC-MS/MS method for simultaneous quantification of ulotaront and its N-desmethyl metabolite in human plasma and application to a clinical study

J Pharm Biomed Anal. 2022 Jan 5:207:114404. doi: 10.1016/j.jpba.2021.114404. Epub 2021 Oct 6.

Abstract

Ulotaront (SEP-363856) is a novel non-D2-receptor-binding agent under development for the treatment of patients with schizophrenia. A highly sensitive liquid chromatography-tandem mass spectrometry (LC-MS/MS) method with lower limit of quantitation of 0.0200 ng/mL (i.e. 20.0 pg/mL) was successfully developed and validated for the simultaneous quantitation of ulotaront and its N-desmethyl metabolite (M11A) in human plasma. Plasma samples were extracted by solid phase extraction with Oasis MCX 96-well plate, followed by a reversed phase LC separation coupled with MS/MS detection in positive mode (m/z 184.1 → 135.0 for ulotaront and 170.1 → 135.0 for M11A). Stable isotope-labeled compounds SEP-363856-d3 and M11A-d4 were used as internal standards (IS) for corresponding analytes. The validated calibration curve range was 0.0200-20.0 ng/mL for both analytes using a 0.200 mL plasma. Extraction recoveries were found to be 75.7% and 75.1% for ulotaront and IS1, and 82.7% and 83.9% for M11A and IS2, respectively. Frozen plasma samples were confirmed to be stable for up to 730 days at both -20 °C and -70 °C. The validated method has been successfully used to evaluate the pharmacokinetics (PK) of ulotaront and M11A in clinical studies. The application to the first-in-human PK study (single ascending dose) presented in this work demonstrated that ulotaront exhibited near dose proportionality for both Cmax (maximum concentration) and AUC (area under the curve) over the dose range from 5 to 125 mg. M11A was found as a minor metabolite with an exposure of about 2-3% of the parent compound.

Keywords: LC-MS/MS; N-desmethyl metabolite; Pharmacokinetics; SEP-856); Single ascending dose; Solid phase extraction; Ulotaront (SEP-363856.

MeSH terms

  • Chromatography, Liquid
  • Humans
  • Plasma*
  • Pyrans
  • Reproducibility of Results
  • Tandem Mass Spectrometry*

Substances

  • Pyrans
  • SEP-363856