Optimized Method for Determination 57 Volatile Organic Compounds in Nitrogen Using GC × GC-FID

J Chromatogr Sci. 2022 Oct 3;60(8):713-724. doi: 10.1093/chromsci/bmab118.

Abstract

The sample containing 57 volatile organic compounds (VOCs) in nitrogen at a nominal 1 ppmv was prepared in our lab using weighting method. A methodology for determination of the 57 VOCs using a two-dimensional gas chromatography equipped with Deans switches and two flame ionization detectors (GC × GC-FID) was developed and validated for resolution, asymmetry, sensitivity, precision (intra-day precision and inter-day precision), linearity, limit of detection (LOD), limit of quantification (LOQ) and accuracy. In this study, resolution, asymmetry and sensitivity of the analytical method were improved,intra-day precisions of all the compounds were <1% and inter-day precisions were between 0.9 and 3.0%. In addition, LOQ and LOD were in the range of 0.024-0.185 ppmv and 0.012-0.092 ppmv, respectively. An excellent linearity was obtained (R2 > 0.9995). At the meantime, the accuracy of the analytical method was evaluated by determining the concentration of a certified reference material.

MeSH terms

  • Chromatography, Gas / methods
  • Flame Ionization / methods
  • Limit of Detection
  • Nitrogen
  • Volatile Organic Compounds* / analysis

Substances

  • Volatile Organic Compounds
  • Nitrogen