Dose reduction potential in cone-beam CT imaging of upper extremity joints with a twin robotic x-ray system

Sci Rep. 2021 Oct 11;11(1):20176. doi: 10.1038/s41598-021-99748-1.

Abstract

Cone-beam computed tomography is a powerful tool for 3D imaging of the appendicular skeleton, facilitating detailed visualization of bone microarchitecture. This study evaluated various combinations of acquisition and reconstruction parameters for the cone-beam CT mode of a twin robotic x-ray system in cadaveric wrist and elbow scans, aiming to define the best possible trade-off between image quality and radiation dose. Images were acquired with different combinations of tube voltage and tube current-time product, resulting in five scan protocols with varying volume CT dose indices: full-dose (FD; 17.4 mGy), low-dose (LD; 4.5 mGy), ultra-low-dose (ULD; 1.15 mGy), modulated low-dose (mLD; 0.6 mGy) and modulated ultra-low-dose (mULD; 0.29 mGy). Each set of projection data was reconstructed with three convolution kernels (very sharp [Ur77], sharp [Br69], intermediate [Br62]). Five radiologists subjectively assessed the image quality of cortical bone, cancellous bone and soft tissue using seven-point scales. Irrespective of the reconstruction kernel, overall image quality of every FD, LD and ULD scan was deemed suitable for diagnostic use in contrast to mLD (very sharp/sharp/intermediate: 60/55/70%) and mULD (0/3/5%). Superior depiction of cortical and cancellous bone was achieved in FDUr77 and LDUr77 examinations (p < 0.001) with LDUr77 scans also providing favorable bone visualization compared to FDBr69 and FDBr62 (p < 0.001). Fleiss' kappa was 0.618 (0.594-0.641; p < 0.001), indicating substantial interrater reliability. In this study, we demonstrate that considerable dose reduction can be realized while maintaining diagnostic image quality in upper extremity joint scans with the cone-beam CT mode of a twin robotic x-ray system. Application of sharper convolution kernels for image reconstruction facilitates superior display of bone microarchitecture.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Cadaver
  • Cone-Beam Computed Tomography / methods*
  • Humans
  • Imaging, Three-Dimensional / methods*
  • Phantoms, Imaging*
  • Radiation Dosage
  • Robotic Surgical Procedures / methods*
  • Upper Extremity / diagnostic imaging*
  • Upper Extremity / surgery
  • X-Rays