Digital workflow and virtual validation of a 3D-printed definitive hollow obturator for a large palatal defect

J Prosthet Dent. 2023 May;129(5):798-804. doi: 10.1016/j.prosdent.2021.08.021. Epub 2021 Oct 9.

Abstract

This clinical report describes how a hollow obturator prosthesis was designed and fabricated for an 82-year-old partially edentulous patient with a large palatal defect. Computer-aided design (CAD) was used to design, articulate, and align the mandibular denture with the obturator prosthesis. The prosthesis was printed, adjusted chairside, rescanned, and made hollow by using a CAD software program. The prosthesis was printed in resin with a dental 3D printer. Quantitative evaluations of clinical (prosthesis dimensions, rest, and occlusal vertical dimensions) and virtual (surface area, volume, weight, interpoint mismatches, spatial overlap) parameters found that the 3D-printed prosthesis required an additional 5% chairside modification. The greatest differences in volume (24.7% less) and weight (22.2% less) were observed when the modified obturator bulb was made hollow via CAD. Hollowing the bulb, therefore, reduced the spatial overlap in volume by 16.8%.