Accuracy of Algorithms and Visual Inspection for Detection of Trigger Asynchrony in Critical Patients : A Systematic Review

Crit Care Res Pract. 2021 Sep 28:2021:6942497. doi: 10.1155/2021/6942497. eCollection 2021.

Abstract

Objective: This study aimed to summarize the accuracy of the different methods for detecting trigger asynchrony at the bedside in mechanically ventilated patients.

Method: A systematic review was conducted from 1990 to 2020 in PubMed, Lilacs, Scopus, and ScienceDirect databases. The reference list of the identified studies, reviews, and meta-analyses was also manually searched for relevant studies. The reference standards were esophageal pressure catheter and/or electrical activity of the diaphragm. Studies were assessed following the QUADAS-2 recommendations, while the review was prepared according to the PRISMA criteria.

Results: One thousand one hundred and eleven studies were selected, and four were eligible for analysis. Esophageal pressure was the predominant reference standard, while visual inspection and algorithms/software comprised index tests. The trigger asynchrony, ineffective expiratory effort, double triggering, and reverse triggering were analyzed. Sensitivity and specificity ranged from 65.2% to 99% and 80% to 100%, respectively. Positive predictive values reached 80.3 to 100%, while the negative predictive values reached 92 to 100%. Accuracy could not be calculated for most studies.

Conclusion: Algorithms/software validated directly or indirectly using reference standards present high sensitivity and specificity, with a diagnostic power similar to visual inspection of experts.

Publication types

  • Review