The predictive role of early recurrences of atrial arrhythmias after pulmonary vein cryoballoon ablation. Is blanking period an outdated concept? Insights from 12-month continuous cardiac monitoring

BMC Cardiovasc Disord. 2021 Oct 7;21(1):483. doi: 10.1186/s12872-021-02300-2.

Abstract

Background: Early recurrences of atrial arrhythmias (ERAA) after atrial fibrillation (AF) catheter ablation do not predict procedural failure. A well-demarcated homogeneous lesion delivered by cryoballoon is less arrhythmogenic, and the recommended three-months blanking period may not refer to cryoballoon ablation (CBA).

Objective: We aimed to evaluate the predictive role of ERAA after second-generation CBA using an implantable loop recorder.

Methods: This prospective observational study enrolled 100 patients (58 males, median age 58) with paroxysmal/persistent AF undergoing pulmonary vein (PV) CBA using second-generation cryoballoon with simultaneous ECG loop recorder implantation. The duration of follow-up was 12 months, with scheduled visits at 3, 6 and 12 months.

Results: 99 patients from 100 completed the 12-month follow-up period. ERAA occurred in 31.3 % of patients. 83.9 % of patients with ERAA also developed late recurrences. The 12-month freedom from AF in patients with ERAA was significantly lower than in those without ERAA (p < 0.0001). Non-paroxysmal AF and longer arrhythmia history were associated with increased risk of both early (HR 3.27; 95 % CI 1.32-8.08; p = 0.010 and HR 1.0147; 95 % CI 1.008-1.086; p = 0.015, respectively) and late recurrences (HR 3.89; 95 % CI 1.67-9.04; p = 0.002 and HR 1.0142; 95 % CI 1.007-1.078; p = 0.019, respectively) of AF. ERAA were another predictor for procedural failure (HR 15.2; 95 % CI (6.42-35.99; p = 0.019).

Conclusions: ERAA occurred in the third of the patients after PV second-generation CBA and are strongly associated with procedural failure. Longer duration of AF history and persistent AF are independent predictors of AF's early and late recurrence.

Publication types

  • Observational Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Action Potentials
  • Atrial Fibrillation / diagnosis
  • Atrial Fibrillation / physiopathology
  • Atrial Fibrillation / surgery*
  • Cryosurgery / adverse effects*
  • Electrocardiography, Ambulatory* / instrumentation
  • Female
  • Heart Rate
  • Humans
  • Male
  • Middle Aged
  • Predictive Value of Tests
  • Prospective Studies
  • Pulmonary Veins / physiopathology
  • Pulmonary Veins / surgery*
  • Recurrence
  • Remote Sensing Technology* / instrumentation
  • Reproducibility of Results
  • Risk Assessment
  • Risk Factors
  • Time Factors
  • Treatment Outcome