Evaluation of Accuracy and Safety of the Next-Generation Up to 180-Day Long-Term Implantable Eversense Continuous Glucose Monitoring System: The PROMISE Study

Diabetes Technol Ther. 2022 Feb;24(2):84-92. doi: 10.1089/dia.2021.0182. Epub 2021 Sep 9.

Abstract

Background: Use of continuous glucose monitoring (CGM) systems is being rapidly adopted as standard of care for insulin-requiring patients with diabetes. The PROMISE study (NCT03808376) evaluated the accuracy and safety of the next-generation implantable Eversense CGM system for up to 180 days. Methods: This was a prospective multicenter study involving 181 subjects with diabetes at 8 USA sites. All subjects were inserted with a primary sensor. Ninety-six subjects had a second sensor, either an identical sensor or a modified sensor (sacrificial boronic acid [SBA]), inserted in their other arm (53 and 43 subjects, respectively). Accuracy was evaluated by comparing CGM to YSI 2300 glucose analyzer (Yellow Springs Instrument [YSI]) values during 10 clinic visits (day 1-180). Confirmed event detection rates, calibration stability, sensor survival, and serious adverse events (SAEs) were evaluated. Results: For primary sensors, the percent CGM readings within 20%/20% of YSI values was 92.9%; overall mean absolute relative difference (MARD) was 9.1%. The confirmed alert detection rate at 70 mg/dL was 93% and at 180 mg/dL was 99%. The median percentage of time for one calibration per day was 56%. Sixty-five percent of the primary sensors survived to 180 days. For the SBA sensors, the percent CGM readings within 20%/20% of YSI values was 93.9%; overall MARD was 8.5%. The confirmed alert detection rate at 70 mg/dL was 94% and at 180 mg/dL was 99%. The median percentage of time for one calibration per day was 63%. Ninety percent of the SBA sensors survived to 180 days. No device- or insertion/removal procedure-related SAEs were reported. Conclusion: These data show the next-generation Eversense CGM system had sustained accuracy and safety up to 180 days, with an improved calibration scheme and survival, using the primary or SBA sensors.

Keywords: Continuous glucose monitoring; Eversense; Implantable sensor; PROMISE study.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Blood Glucose
  • Blood Glucose Self-Monitoring
  • Diabetes Mellitus, Type 1* / drug therapy
  • Diabetes Mellitus, Type 2*
  • Humans
  • Prospective Studies
  • Reproducibility of Results

Substances

  • Blood Glucose