Effect of spatial averaging on the amplitude ring ratio in multifocal electroretinography

Doc Ophthalmol. 2022 Feb;144(1):41-52. doi: 10.1007/s10633-021-09850-y. Epub 2021 Sep 10.

Abstract

Purpose: The purpose of this study was to evaluate the effect of spatial averaging on the multifocal electroretinography (mfERG) amplitude ring ratios used in screening for hydroxychloroquine (HCQ) toxicity.

Methods: This was a retrospective review of the records of patients screened for HCQ retinopathy at the USF Eye Institute (University of South Florida) during the period of 2015-2020. Patients were tested binocularly with Diagnosys mfERG system (Diagnosys LLC, Lowell, MA). Only the records of patients referred internally were used. The effects of the lowest level (level 1, or 8%) of spatial averaging on the P1 amplitude ring ratios used for screening of HCQ maculopathy: R1/R2, R2/R5, R5/R3 and R5/R4, were evaluated.

Results: The records of 40 patients (4 males, 36 females) aged 54.4 ± 14.1 years were selected for analysis. The use of spatial averaging had a significant effect on P1 amplitudes, and on the ring ratios and this effect was correlated with the magnitude of the amplitudes and the ratios. Spatial averaging diminished P1 amplitude significantly in ring 1 (p < 0.0001) and increased it slightly in ring 4 (p < 0.05), while it had no effect on the amplitude of the other three rings. Although as a group spatial averaging had a moderate effect on the R1/R2 ratio (~ -15%), on an individual basis the range was wide, from -36 to 43%. The effect on the other ring ratios was similar: The average group effect was ~ -5%, ~ -3.4% and ~ -4% for R2/R5, R5/R3 and R5/R4 ratios, but individual effects ranged from 0.18% to -27.3%, 0.9% to -14.2% and 0.9% to -26.2%, respectively.

Conclusions: For all ring ratios used in this analysis, spatial averaging has a substantial effect on the ring ratio, which could affect the interpretation of the results. Therefore, use of spatial averaging should be avoided when analyzing mfERG results for HCQ screening.

Keywords: Hydroxychloroquine; Maculopathy; Multifocal ERG; Ring ratios.

MeSH terms

  • Electroretinography / methods
  • Female
  • Humans
  • Hydroxychloroquine / adverse effects
  • Macular Degeneration*
  • Male
  • Retina
  • Retinal Diseases* / chemically induced
  • Retinal Diseases* / diagnosis

Substances

  • Hydroxychloroquine