The aim of the study was to evaluate the clinical performance of FTD SARS-CoV-2 compared to the RealStar RT-PCR kit 1.0. The analysis of 100 nasopharyngeal swabs showed an overall agreement of 88 %. The positive percentage agreement was 85.6 % and the negative percentage agreement was 91 %. In conclusion we observed a substantial agreement among the two methods, with discrepancies mainly observed in specimens with relatively low amount of viral RNA.
Keywords: Covid-19; Ct values; FTD SARS-CoV-2; RT-PCR; RealStar RT-PCR; SARS-CoV-2.