Biomechanical Analysis of the Kirschner-Wire Depth of the Modified Tension Band Wiring Technique in Transverse Patellar Fractures: An Experimental Study Using the Finite-Element Method

Clin Orthop Surg. 2021 Sep;13(3):315-319. doi: 10.4055/cios20253. Epub 2021 Jul 16.

Abstract

Backgroud: Modified tension band wiring is one of the most preferred surgical methods for transverse patellar fractures. However, the optimal depth or sagittal position of a Kirschner wire (K-wire) in modified tension band wiring has yet to be determined. The purpose of this study was to evaluate whether the depth of a K-wire affects the biomechanical characteristics of modified tension band wiring using the finite-element method.

Methods: A patella model was designed with a cuboid shape (length, 34.3 mm; width, 44.8 mm; and thickness, 22.4 mm) and divided into the cortical and cancellous bone parts. A transverse fracture line was formed on the midline of the cuboid shape model. The cuboidal model was applied to modified tension band wiring. The depth or sagittal position of the K-wire was divided into superficial, center, and deep. With the Abaqus v2017 program (Dassault System Inc.), the distal part of the model was fixed, and a tensile load of 850 N was applied to the proximal part of the model at an angle of 45°. The maximum pressures of the cortical and cancellous bones at the fracture plane were measured. The largest von Mises values of the K-wire and stainless steel wire were also measured. The fracture gap on the distracted or anterior side was measured.

Results: In deep K-wire placement, the highest peak von Mises values of the cortical and cancellous bones were observed. The K-wire and stainless steel wire showed the highest von Mises values in deep K-wire placement. The fracture gap was also largest in deep K-wire placement.

Conclusions: The depth of the K-wire affects the biomechanical characteristics of modified tension band wiring. Deep placement of the K-wire will be more favorable for bone union than the empirically known 5-mm anterior or center placement of the K-wire.

Keywords: Patella; Tension band wiring; Transverse fracture.

MeSH terms

  • Biomechanical Phenomena
  • Bone Wires*
  • Finite Element Analysis
  • Fracture Fixation, Internal / methods*
  • Fractures, Bone / surgery*
  • Humans
  • Patella / injuries*
  • Patella / surgery*