Implantable vagus nerve stimulation system performance is not affected by internal or external defibrillation shocks

J Interv Card Electrophysiol. 2022 Apr;63(3):555-560. doi: 10.1007/s10840-021-01050-7. Epub 2021 Aug 31.

Abstract

Purpose: Autonomic regulation therapy (ART) for heart failure (HF) is delivered using vagus nerve stimulation (VNS), and has been associated with improvement in cardiac function and HF symptoms. VNS is delivered using an implantable pulse generator (IPG) and a lead placed around the cervical vagus nerve. Because HF patients may receive concomitant cardiac defibrillation therapy, testing was conducted to determine the effect of defibrillation (DF) on VNS system performance.

Methods: Normal swine (n = 4) with VNS system implants on the right cervical vagus nerve received sequential defibrillation shocks with three defibrillation systems: an implantable cardioverter defibrillator (ICD), a subcutaneous ICD (S-ICD), and an external cardioverter defibrillator (ECD). Each system delivered a series of bipolar high-energy shocks and reverse-polarity high-energy shocks.

Results: The specified cardiac defibrillation shocks were delivered successfully from each of the three defibrillation systems to all animals. After each shock series, interrogation of the IPG confirmed that software and data were unchanged from pre-programmed values. After all of the defibrillation shocks were delivered, the IPGs underwent and passed comprehensive electrical testing demonstrating proper system function. No shifts in IPG parameters or ART system failures were observed, and histologic evaluation of the vagus nerve revealed no anatomic changes.

Conclusions: Implantable VNS systems were tested in vivo for immunity to defibrillation via ICD, S-ICD, and ECD, and were found to be unaffected by a series of high-energy defibrillation shocks. These results confirm that ART systems are capable of continuing to function after defibrillation and the cervical vagus nerve is anatomically unaffected.

Keywords: Autonomic nervous system; Autonomic regulation therapy; Defibrillation; Heart failure; Implantable cardioverter defibrillator; Vagus nerve; Vagus nerve stimulation.

MeSH terms

  • Animals
  • Anti-Arrhythmia Agents
  • Defibrillators, Implantable* / adverse effects
  • Electric Countershock / methods
  • Electrocardiography
  • Heart Failure* / therapy
  • Humans
  • Swine
  • Vagus Nerve Stimulation*

Substances

  • Anti-Arrhythmia Agents