Quantitative time-of-flight MR angiography for simultaneous luminal and hemodynamic evaluation of the intracranial arteries

Magn Reson Med. 2022 Jan;87(1):150-162. doi: 10.1002/mrm.28969. Epub 2021 Aug 10.

Abstract

Purpose: To report a quantitative time-of-flight (qTOF) MRA technique for simultaneous luminal and hemodynamic evaluation of the intracranial arteries.

Methods: Implemented using a thin overlapping slab 3D stack-of-stars based 3-echo FLASH readout, qTOF was tested in a flow phantom and for imaging the intracranial arteries of 10 human subjects at 3 Tesla. Display of the intracranial arteries with qTOF was compared to resolution-matched and scan time-matched standard Cartesian 3D time-of-flight (TOF) MRA, whereas quantification of mean blood flow velocity with qTOF, done using a computer vision-based inter-echo image analysis procedure, was compared to 3D phase contrast MRA. Arterial-to-background contrast-to-noise ratio was measured, and intraclass correlation coefficient was used to evaluate agreement of flow velocities.

Results: For resolution-matched protocols of similar scan time, qTOF portrayed the intracranial arteries with good morphological correlation with standard Cartesian TOF, and both techniques provided superior contrast-to-noise ratio and arterial delineation compared to phase contrast (20.6 ± 3.0 and 37.8 ± 8.7 vs. 11.5 ± 2.2, P < .001, both comparisons). With respect to phase contrast, qTOF showed excellent agreement for measuring mean flow velocity in the flow phantom (intraclass correlation coefficient = 0.981, P < .001) and good agreement in the intracranial arteries (intraclass correlation coefficient = 0.700, P < .001). Stack-of-stars data sampling used with qTOF eliminated oblique in-plane flow misregistration artifacts that were seen with standard Cartesian TOF.

Conclusion: qTOF is a new 3D MRA technique for simultaneous luminal and hemodynamic evaluation of the intracranial arteries that provides significantly greater contrast-to-noise ratio efficiency than phase contrast and eliminates misregistration artifacts from oblique in-plane blood flow that occur with standard 3D TOF.

Keywords: flow; intracranial; phase contrast; qTOF; quantitative; time-of-flight.