Validation of ambulatory monitoring devices to measure energy expenditure and heart rate in a military setting

Physiol Meas. 2021 Aug 27;42(8). doi: 10.1088/1361-6579/ac19f9.

Abstract

Objectives.To investigate the validity of different devices and algorithms used in military organizations worldwide to assess physical activity energy expenditure (PAEE) and heart rate (HR) among soldiers.Design.Device validation study.Methods. Twenty-three male participants serving their mandatory military service accomplished, firstly, nine different military specific activities indoors, and secondly, a normal military routine outdoors. Participants wore simultaneously an ActiHeart, Everion, MetaMax 3B, Garmin Fenix 3, Hidalgo EQ02, and PADIS 2.0 system. The PAEE and HR data of each system were compared to the criterion measures MetaMax 3B and Hidalgo EQ02, respectively.Results. Overall, the recorded systematic errors in PAEE estimation ranged from 0.1 (±1.8) kcal.min-1to -1.7 (±1.8) kcal.min-1for the systems PADIS 2.0 and Hidalgo EQ02 running the Royal Dutch Army algorithm, respectively, and in the HR assessment ranged from -0.1 (±2.1) b.min-1to 0.8 (±3.0) b.min-1for the PADIS 2.0 and ActiHeart systems, respectively. The mean absolute percentage error (MAPE) in PAEE estimation ranged from 29.9% to 75.1%, with only the Everion system showing an overall MAPE <30%, but all investigated devices reported overall MAPE <1.4% in the HR assessment.Conclusions. The present study demonstrated poor to moderate validity in terms of PAEE estimation, but excellent validity in all investigated devices in terms of HR assessment. Overall, the Everion performed among the best in both parameters and with a device placement on the upper arm, the Everion system is particularly useful during military service, as it does not interfere with other relevant equipment.

Keywords: fitness trackers; military personnel; physiological monitoring; wearable electronic devices.

MeSH terms

  • Energy Metabolism
  • Fitness Trackers
  • Heart Rate
  • Humans
  • Male
  • Military Personnel*
  • Monitoring, Ambulatory