A Comparative Study of the Accuracy of Dental CAD Programs in Designing a Fixed Partial Denture

J Prosthodont. 2022 Mar;31(3):215-220. doi: 10.1111/jopr.13406. Epub 2021 Aug 9.

Abstract

Purpose: Error testing at each stage of prosthetic manufacturing remains relatively underdeveloped for computer-aided design/computer-aided manufacturing methods, and no experimental studies have validated the computer-aided design programs. This study aimed to test the accuracy and trueness of the computer-aided design of a three-unit fixed prosthesis.

Materials and methods: Three computer-aided design programs (Exocad, Dental System™, and inLab 16) were tested on the designs of a three-unit fixed partial denture, and a three-dimensional analysis program was used to calculate the internal clearance error for the computer-aided design prostheses. The Kruskal-Wallis and Dunn's post hoc tests were used to reveal significant differences in trueness between the three computer-aided design programs (α < 0.05).

Results: Dental System™ showed the lowest mean error values for #24 and #26 at the mesial margin (both 0 µm), mesial wall (0.10, 0.12 µm, respectively), occlusal surface (-0.05, 0.10 µm), distal wall (0.23, -0.02 µm), and distal margin (both 0 µm). In sum, except for the mesial margin and distal margin site of tooth #26, the mean error value of Dental System™ was statistically the lowest, followed by those of Exocad and inLab 16 (p < 0.003).

Conclusions: The accuracy of computer-aided design differed according to the type of computer-aided design program. Dental System™ achieved the best trueness at the margins, axial walls, and occlusal surface, followed by Exocad and inLab 16.

Keywords: Computer-aided design; dental marginal adaptation; dimensional measurement accuracy; prosthesis design.