Accelerate gas diffusion-weighted MRI for lung morphometry with deep learning

Eur Radiol. 2022 Jan;32(1):702-713. doi: 10.1007/s00330-021-08126-y. Epub 2021 Jul 13.

Abstract

Objectives: Multiple b-value gas diffusion-weighted MRI (DW-MRI) enables non-invasive and quantitative assessment of lung morphometry, but its long acquisition time is not well-tolerated by patients. We aimed to accelerate multiple b-value gas DW-MRI for lung morphometry using deep learning.

Methods: A deep cascade of residual dense network (DC-RDN) was developed to reconstruct high-quality DW images from highly undersampled k-space data. Hyperpolarized 129Xe lung ventilation images were acquired from 101 participants and were retrospectively collected to generate synthetic DW-MRI data to train the DC-RDN. Afterwards, the performance of the DC-RDN was evaluated on retrospectively and prospectively undersampled multiple b-value 129Xe MRI datasets.

Results: Each slice with size of 64 × 64 × 5 could be reconstructed within 7.2 ms. For the retrospective test data, the DC-RDN showed significant improvement on all quantitative metrics compared with the conventional reconstruction methods (p < 0.05). The apparent diffusion coefficient (ADC) and morphometry parameters were not significantly different between the fully sampled and DC-RDN reconstructed images (p > 0.05). For the prospectively accelerated acquisition, the required breath-holding time was reduced from 17.8 to 4.7 s with an acceleration factor of 4. Meanwhile, the prospectively reconstructed results showed good agreement with the fully sampled images, with a mean difference of -0.72% and -0.74% regarding global mean ADC and mean linear intercept (Lm) values.

Conclusions: DC-RDN is effective in accelerating multiple b-value gas DW-MRI while maintaining accurate estimation of lung microstructural morphometry, facilitating the clinical potential of studying lung diseases with hyperpolarized DW-MRI.

Key points: • The deep cascade of residual dense network allowed fast and high-quality reconstruction of multiple b-value gas diffusion-weighted MRI at an acceleration factor of 4. • The apparent diffusion coefficient and morphometry parameters were not significantly different between the fully sampled images and the reconstructed results (p > 0.05). • The required breath-holding time was reduced from 17.8 to 4.7 s and each slice with size of 64 × 64 × 5 could be reconstructed within 7.2 ms.

Keywords: Deep learning; Diffusion magnetic resonance imaging; Lung.

MeSH terms

  • Deep Learning*
  • Diffusion Magnetic Resonance Imaging
  • Humans
  • Lung / diagnostic imaging
  • Magnetic Resonance Imaging
  • Pulmonary Disease, Chronic Obstructive*
  • Retrospective Studies
  • Xenon Isotopes

Substances

  • Xenon Isotopes